EC/IR II Measurement Limits: Uncertainty, Mouth Alcohol, Breath–Blood Comparisons, and Timing
Evaluate EC/IR II uncertainty, breath–blood comparisons, timing research, mouth alcohol, and interferents without treating study results as automatic corrections.



A credible analysis identifies both what a measurement supports and what it leaves unresolved. The EC/IR II’s design features, a passing check, and a reported concentration are all relevant. So are the conditions of sampling, the scope of validation, the reporting convention, and the assumptions behind an inference about blood alcohol or an earlier time.
This chapter reviews selected instrument-related research and explains how to use it responsibly. A study is not an automatic numerical correction for every ticket. The question is whether its methods, population, instrument configuration, and uncertainty fit the proposition being offered in the case.
1. Accuracy, precision, bias, and uncertainty
Accuracy concerns closeness to an appropriate reference. Precision concerns agreement among repeated measurements. Bias describes a systematic difference under specified conditions. Uncertainty characterizes the range reasonably associated with a measurement under a defined evaluation. These concepts answer related but different questions.
Two close results can show agreement while sharing a common source of bias. A reference check may support accuracy under its test conditions while leaving a specimen-origin question unresolved. A study’s average bias does not tell you the error in one individual result. Use the terms carefully so that a conclusion about one property is not silently converted into a conclusion about another.
An expert should identify the measurement being evaluated, the sources of uncertainty included, the confidence or coverage convention, and what was excluded. A numerical uncertainty statement without those details can appear more informative than it is.
2. A Wisconsin EC/IR II uncertainty study
Brockley-Drinkman and Barkholtz’s study, published in the Journal of Analytical Toxicology in 2020, developed a bottom-up uncertainty estimate using Wisconsin EC/IR II data from 2014–2018. It reported expanded uncertainty of 0.00458 g/210 L at 95% confidence within its specified analysis. DOI 10.1093/jat/bkz050; PubMed record.
The appropriate question is whether the underlying data, assumptions, controls, and reporting approach apply to the unit and proposition at issue. It is not appropriate to subtract 0.00458 from every NC result as though the article created a statutory deduction. Nor does a 95% uncertainty statement by itself give the probability that a particular defendant’s true concentration was below a legal threshold.
Ask an expert to explain the difference between uncertainty in an instrument measurement and uncertainty in an inference about a person at another time. Those are different analytical tasks and may require different inputs.
3. Paired breath and blood measurements
Skaggs and colleagues compared EC/IR II breath results with blood analysis in a controlled drinking study involving 114 volunteers and a 30-minute deprivation period. Their untruncated comparison reported a mean difference of 0.00754 and 95% limits of agreement of −0.00705 to 0.0221, for blood minus breath, expressed in g/100 mL under the paper’s comparison convention. The paper also evaluated Wisconsin truncation. JAT 46(6), 683–688, DOI 10.1093/jat/bkab086; PubMed.

Limits of agreement describe the observed comparison under the study’s method; they are not a confidence interval around this defendant’s result. The graph below is a study-summary graphic, not a distribution of all NC tests or an instrument failure rate. Wisconsin reporting assumptions should not be imported into North Carolina.
If a case contains both breath and blood measurements, examine collection times, specimen handling, analytical methods, units, and reporting conventions before calling the results contradictory. Two specimens taken at different times may answer different questions. A scientifically useful comparison needs the chronology and the method, not merely two numbers placed side by side.
4. Timing research published in 2026
O’Connor, Wang, and Vargas studied 42 adults under supervised social-drinking conditions, using EC/IR II measurements at approximately 30-minute intervals. Time from the last drink to observed peak ranged from −23 to 76 minutes, with a median of 13 and mean of 17. The design allowed varied drinking duration and food intake. JAT 50(6), bkag038, DOI 10.1093/jat/bkag038; full-text access through PubMed.

This study illustrates variability under its conditions. Its sample size and measurement intervals matter. An observed peak based on intermittent measurements is not an exact continuous record of every participant’s concentration. The findings do not create a universal waiting time, a personal sobriety calculator, or a reliable estimate for someone whose drinking history is unknown.
For a case-specific timing opinion, ask what is known about drinking times, amounts, food, specimen times, and other relevant circumstances. Separate measured facts from reported history and assumptions. Test alternative assumptions rather than hiding them inside a single confident-looking estimate. Remember that the wording of North Carolina’s charged offense matters when evaluating a timing argument; Chapter 7 connects the science to the legal theory.
5. Breath physiology and the limits of simple conversion
Research by Hlastala and colleagues examines alcohol exchange in the airways and the effects of breathing pattern and lung size. These publications are useful for understanding why breath–blood relationships involve physiology. They are not, by themselves, findings that a particular EC/IR II unit malfunctioned. Hlastala, 2010, DOI 10.1111/j.1556-4029.2009.01269.x; Hlastala and Anderson, 2007, DOI 10.1007/s10439-006-9216-3.
A legal presentation should specify whether the expert is questioning a conversion to a blood concentration, the quality of a collected specimen, or an inference about the time of driving. Those propositions require different support. North Carolina’s statutory breath unit is itself part of the legal framework; an abstract physiological observation does not eliminate that framework.
This discussion is not a set of techniques for changing a test result. The useful application is retrospective evaluation by a qualified expert with an identified scientific question and the actual record.
6. Mouth alcohol and environmental concerns
The supplied WV manual addresses mouth-alcohol messages, high blanks, and alcohol-containing products in the testing environment. It recommends a clean testing location and identifies conditions that warrant attention. WV manual, pages 3 and 16–18. North Carolina’s observation requirements and the actual instrument’s safeguards must be evaluated independently.
Distinguish a possible mechanism from a documented event. A history of reflux does not establish that regurgitation occurred during this test. A product containing alcohol does not establish exposure at a concentration and time capable of affecting the accepted sample. A high-blank message before a successful test does not, by itself, establish that the later test was affected, either. Each proposition needs evidence about timing, conditions, instrument response, and the corrective sequence.
An expert may be able to explain why a mechanism is plausible while being unable to quantify its effect in a case. That limitation should be visible. A qualified, bounded opinion is more informative than a speculative claim presented as a measured correction.
7. Interferents and the importance of test mode
Karen Chan’s EC/IR II research examined isopropanol and ethanol mixtures and distinguished subject testing with interference detection from another testing mode. The study illustrates why configuration and enabled safeguards matter when applying research to a case. Canadian Society of Forensic Science Journal 51(3–4), 67–82, DOI 10.1080/00085030.2018.1491691.
Do not translate “a substance affected a tested condition” into “this substance explains this ticket.” Obtain the complete paper before relying on detailed thresholds or methods. Compare the configuration and establish the relevant exposure. The same discipline applies to broad manufacturer claims of selectivity: understand what was tested and what the statement covers.
8. A research-reading checklist
| Question | Why it changes the interpretation |
|---|---|
| Was the study about this model and configuration? | Similar names do not establish equivalent measurement behavior. |
| Was it a controlled laboratory test, volunteer study, model, or field dataset? | Each design answers a different type of question. |
| What was the reference method? | “Error” requires a defined comparison. |
| What were the sampling and timing conditions? | Results may depend on conditions absent from the case. |
| What does the reported interval mean? | Agreement limits, uncertainty intervals, and confidence intervals differ. |
| What did the authors exclude or leave unresolved? | The boundary of a conclusion is part of the conclusion. |
| Does the opinion fit the actual legal proposition? | A scientific observation may be real but irrelevant to the disputed issue. |
Use research to improve the question, then use case-specific evidence to evaluate the answer. The discovery chapter turns that approach into a practical work plan.
Questions readers ask
Can the Wisconsin uncertainty figure simply be subtracted from every North Carolina breath result?
No. The study’s figure belongs to its defined data, assumptions, controls, and reporting analysis. Applying it to another event requires a justified comparison. The paper does not create an automatic statutory deduction from a North Carolina ticket.
Are a study’s limits of agreement a confidence interval around one person’s result?
No. The chapter distinguishes agreement limits in a paired study from an uncertainty estimate or a case-specific confidence statement. Examine the study’s reference method, timing, units, population, and reporting conventions before applying its findings.
Can the 2026 timing study tell me exactly when a particular person peaked?
No. It illustrates variability under its study conditions, with a limited sample and intermittent measurements. A case-specific timing opinion requires the actual collection times, supported history, relevant assumptions, and an explanation of the limits of the calculation.
Places behind the legal system
Regional context photographs. These images do not depict the people or events in any case discussed here.


Explore the North Carolina courthouse atlas and county map →
Continue exploring
Previous: Reading an EC/IR II Test Ticket: Blanks, Standards, Samples, Errors, and Missing Records
Read the related guide and its sources.
Explore →Next: Auditing an EC/IR II DWI Case: Discovery, Preservation, Timelines, and Expert Questions
Read the related guide and its sources.
Explore →Master guide: Intox EC/IR II
Read the related guide and its sources.
Explore →Sources and further research
Follow the original authority and the cited pinpoint. The list distinguishes source types; it is not an exhaustive citator review.
- ResearchDOI 10.1093/jat/bkz050DOI 10.1093/jat/bkz050. Discussed in: A Wisconsin EC/IR II uncertainty study.
- ResearchBrockley-Drinkman and Barkholtz — Wisconsin evidential breath uncertainty (PubMed)PubMed record. Discussed in: A Wisconsin EC/IR II uncertainty study.
- ResearchJAT 46(6), 683–688, DOI 10.1093/jat/bkab086JAT 46(6), 683–688, DOI 10.1093/jat/bkab086. Discussed in: Paired breath and blood measurements.
- ResearchSkaggs and colleagues — controlled breath–blood comparison (PubMed)PubMed. Discussed in: Paired breath and blood measurements.
- ResearchJAT 50(6), bkag038, DOI 10.1093/jat/bkag038JAT 50(6), bkag038, DOI 10.1093/jat/bkag038. Discussed in: Timing research published in 2026.
- ResearchO’Connor, Wang, and Vargas — 2026 breath-concentration timing study (PubMed)full-text access through PubMed. Discussed in: Timing research published in 2026.
- ResearchHlastala, 2010, DOI 10.1111/j.1556-4029.2009.01269.xHlastala, 2010, DOI 10.1111/j.1556-4029.2009.01269.x. Discussed in: Breath physiology and the limits of simple conversion.
- ResearchHlastala and Anderson, 2007, DOI 10.1007/s10439-006-9216-3Hlastala and Anderson, 2007, DOI 10.1007/s10439-006-9216-3. Discussed in: Breath physiology and the limits of simple conversion.
- ResearchCanadian Society of Forensic Science Journal 51(3–4), 67–82, DOI 10.1080/00085030.2018.1491691Canadian Society of Forensic Science Journal 51(3–4), 67–82, DOI 10.1080/00085030.2018.1491691. Discussed in: Interferents and the importance of test mode.